Digi-Sense RTD-Sonde aus Edelstahl
RTD Bendable-Seath Probes are designed to measure temperature in a confined space.
Marke: Digi-Sense 08117-97
This item is not returnable.
View return policy
Description
- The first 3 in. of the probe tip are not to be bent.
- Time Constant: 3 seconds or greater
- Custom designs or changes to this product are available upon request.
- Maximum error for probe assembly is ±1.1°C + 0.12% of reading below 300°F, ±0.35% of reading above 300°F.
- Alpha coefficient is 0.003850 omega/omega/°C (DIN IEC 751).
- Resistance at ice point (0°C) is 100 omega.
- Note: Time constant is the time required to reach 63.2% of the new reading.
- To determine how long it will take to reach 99% of the new reading multiply the time constant by five.
- Note: the temperature range is for the probe tip only.
- The handle material may be affected if the probe is rated higher than the handle depending on exposure time, application, and handle material.
- ABS handles are rated to 220°F (104°C)
- PVC handles are rated to 200°F (93°C)
- Stainless Steel handles are rated at 450°F (232°C)
- Mini Connectors are rated to 392°F (200°C)
- The handle material may be affected if the probe is rated higher than the handle depending on exposure time, application, and handle material.
Specifications
224,6mm | |
8.8 in. | |
-50 °C | |
-50°bis +500 °C |
RTD-Sonde | |
500°C | |
-58°F to 932°F |